64 Digitizers Command Reference Chapter 3
DIAGnostic:DAC:OFFSet
DIAGnostic:DAC:OFFSet[<channel>] <voltage> writes the specified voltage
value to the calibration offset DAC of the specified channel when the
DAC:GAIN command is sent. This offset voltage value is not used unless
a DAC:GAIN <value> is sent to the calibration gain DAC. This command is
a factory diagnostic routine.
Parameters
Comments Executable when initiated: NO
Coupled Command: NO
Reset (*RST) Condition: None
DIAGnostic:DAC:OFFSet:RAMP
DIAGnostic:DAC:OFFSet[<channel>]:RAMP <count> outputs to the
specified channel, a ramp of DAC values from 0 to 255 with the DAC code
changing approximately every 100 msec. This command is a factory
diagnostic routine.
Parameters
Comments Using the <count> Parameter: The <count> parameter defines the number
of ramps to output. Approximately 37.35 full ramps are output each second.
A count of 2240 will output ramps for approximately 60 seconds.
Executable when initiated: NO
Coupled Command: NO
Reset (*RST) Condition: None
Name Type Range of Values Default Value
voltage numeric -2.5 to +2.5 none
Name Type Range of Values Default Value
count numeric 1 to 32767 none