Agilent Technologies E1564A Stereo Receiver User Manual


 
112 Digitizers Command Reference Chapter 3
Executable when initiated: YES
Coupled command: NO
Related commands: *OPC?, *WAI
Reset (*RST) condition: none
*OPC?
*OPC? causes the E1563A and E1564A Digitizers to wait for all pending
operations to complete. A single ASCII “1” is then placed in the output
queue.
Comments INIT vs. OPC?: The INIT command is considered complete when the
measurement is started. *OPC? will return “1” once INIT is processed and
measurements start but the instrument may not be finished taking all
readings initiated.
Executable when initiated: YES
Coupled command: NO
Related commands: *OPC?, *WAI
Reset (*RST) condition: none
*RST
*RST resets the E1563A and E1564A Digitizers as follows:
Sets all commands to their *RST state
Aborts a calibration (CAL:STATe ON)
Resets the CAL:STATe to OFF
Aborts all pending operations.
*RST does not affect:
The output queue
The Service Request and Standard Event Status Enable Registers
The enable unmasks for the QUEStionable Status Registers
Calibration data
Comments Executable when initiated: Yes
Coupled command: No
Reset (*RST) condition: none