Agilent Technologies E1564A Stereo Receiver User Manual


 
Digitizers Command Reference 83Chapter 3
Parameters
Comments Sample Slopes and Periods: A rising or falling edge for the sample slope
can be specified if the source is set to EXTernal (see SAMPle:SLOPe). A
sampling period can be specified if the sample source is set to TIMer (see
SAMPle:TIMer).
Slave Mode: TRIG:MODE SLAVe<n> forces the sample source to be the
appropriate TTL trigger line. Attempts to change the sample source while
TRIG:MODE is SLAVe<n> will result in a settings conflict error message.
Executable when initiated: NO
Coupled command: YES. TRIG:MODE SLAVe<n> forces a specified TTL
trigger line to the sample source. A settings conflict occurs if you attempt
to change this dedicated line with SAMPle:SOURce. TTL sources may
conflict with the output subsystem. Specifying a TTL source will force the
output to be disabled. See the OUTPut subsystem.
Reset (*RST) condition: TIMer source with 0.0000013 second sampling
interval per reading.
SAMPle:SOURce?
SAMPle:SOURce? queries the present source setting for the sample signal.
The returned string is HOLD, TIMer, TTLT0-7 or EXT.
Name Type Point of Source Default
HOLD discrete SAMPle[:IMMediate] none
TIMer discrete Uses specified SAMPle:TIMer
<interval> as sample rate
none
TTLT0-7 discrete VXIbus TTL trigger lines none
EXTernal discrete “Sample” pin on D-sub connector none