Agilent Technologies E1564A Stereo Receiver User Manual


 
16 Configuring the Digitizer Modules Chapter 1
Figure 1-2. E1564A 2-Channel Digitizer Front Panel
plug&play
Front Panel Indicators
Failed LED
: Illuminates momentarily during digitizer power-on.
Access LED: Illuminates when the backplane is communicating with the
digitizer.
Error LED: Illuminates only when an error is present in the digitizer’s driver
error queue. The error can result from improperly executing a command
or the digitizer being unable to pass self-test or calibration.
Sample LED: Illuminates while the digitizer samples the input for a
measurement. Typically blinks for slow sample rates and is on
steady-state for high sample rates.
User Input Terminals
The E1564A Digitizer front panel contains four female connectors for user
inputs. Mating male connectors are supplied with the module. However,
the user must provide the input cable and connect the male connector to
the cable. See "User Cabling Considerations" for connecting
user-supplied cables.
External Trigger Input/Calibration Bus Output
The front panel contains a 9-pin D-subminiature connector for external
(TTL) trigger inputs and for calibration bus outputs. The E1564A
4-Channel Digitizer has a calibration bus output (High, Low and Guard)
and a programmable short. The user must provide the the appropriate
cable to the external trigger input/calibration bus output.