Agilent Technologies E1564A Stereo Receiver User Manual


 
66 Digitizers Command Reference Chapter 3
DIAGnostic:INTerrupt:LINE
DIAGnostic:INTerrupt:LINE 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 sets the interrupt line to be
used. Specifying the “0” parameter disables all interrupts.
NOTE The STATus subsystem will not work if interrupts are disabled (STATus:
OPERation and STATus:QUEStionable). Use DIAG:STATus? to disable
interrupts.
Comments Power-On Setting: Power-on default setting is interrupt line “1”.
Executable when initiated: NO
Coupled Command: NO
Reset (*RST) Condition: Interrupt line setting is unchanged
DIAGnostic:INTerrupt:LINE?
DIAGnostic:INTerrupt:LINE? queries the interrupt line setting. Returns a
number “0” through “7” to indicate interrupt line 1 through 7. A “0” returned
indicates all interrupts are disabled.
NOTE The STATus subsystem will not work if interrupts are disabled (STATus:
OPEReration and STATus:QUEStionable). Use DIAG:STATus? to disable
interrupts.
DIAGnostic:MEMory:SIZE
DIAGnostic:MEMory:SIZE <size> sets the memory size value in calibration
memory. Your module comes standard with 4 Mbytes of RAM. You can
replace this with PC SIMM modules of up to 128 Mbytes. See Chapter 1
for the procedure for adding RAM to your module.
NOTE This command is required and used only when you change the size of RAM
on the module. You then use this command to set the new memory size
value in calibration memory.