Agilent Technologies E1564A Stereo Receiver User Manual


 
160 Digitizers Verification Tests Appendix D
Date___________________
PERFORMANCE TEST RECORD
H
HH
H E1563A 2-Channel Digitizer H
HH
H E1564A 4-Channel Digitizer
CHANNEL: H
HH
H1 H
HH
H2 H
HH
H3 H
HH
H4
Test
Input
Digitizer
Range
Low
Limit
Measured
Reading
High
Limit
Meas
Uncert
Test
Accuracy
Ratio
Zero Offset Test
0 62 mV -.000020 .000020 N/A N/A
0 250 mV -.000078 .000078 N/A N/A
0 1V -.000300 .000300 N/A N/A
0 4V -.001200 .001200 N/A N/A
0 16V -.021000 .021000 N/A N/A
0 64V -.028000 .028000 N/A N/A
0 256V -.079000 .079000 N/A N/A
Noise Test
0 62 mV 0 57 mV max N/A N/A
0 250 mV 0 180 mV max N/A N/A
0 1V 0 720 mV max N/A N/A
0 4V 0 2.88 mV max N/A N/A
0 16V 0 14.7 mV N/A N/A
0 64V 0 48 mV N/A N/A
0 256V 0 189 mV N/A N/A
Gain Test
55 mV 62 mV 54.9613 mV 55.0387 mV .0000011 >10:1
-55 mV 62 mV -55.0387 mV -54.9613 mV .0000011 >10:1
240 mV 250 mV 239.84 mV 240.16 mV .0000022 >10:1
-240 mV 250 mV -240.16 mV -239.84 mV .0000022 >10:1
0.95V 1V 0.949377 V 0.950623 V .0000069 >10:1
-0.95V 1V -0.950623 V -0.949377 V .0000069 >10:1
3.8V 4V 3.79751 V 3.80249 V .000023 >10:1
-3.8V 4V -3.80249 V -3.79751 V .000023 >10:1