52 Digitizers Command Reference Chapter 3
Parameters
Comments Maximum Allowed Values: The maximum allowed <value> depends on the
range setting. An error will occur if you try to set a level that exceeds the
range setting. Changing the range after setting the limit value will change the
limit value. The percent of full scale is kept constant. Allowable maximum
values for the upper limit by range and the associated resolution follow.
Executable when initiated: NO
Coupled Command: YES. Range changes will change the value. The
percent of full scale of the range will be kept constant. For example, on
the 4 volt range (with a 2V limit) a range change to 16V will set a new limit
of 8V.
Reset (*RST) Condition: +252 Volts
CALCulate:LIMit:UPPer:DATA?
CALCulate[<channel>]:LIMit:UPPer:DATA? [MIN | MAX] queries the upper
limit value set for the specified channel.
CALCulate:LIMit:UPPer[:STATe]
CALCulate[<channel>]:LIMit:UPPer[:STATe] OFF | 0 | ON | 1 enables the
upper limit checking for the specified channel. Use LIMit:UPPer:
DATA <value> to set the actual limit value to be tested against.
Comments Executable when initiated: YES
Coupled command: YES. Setting the upper state ON will cause
LIMit:LOWer[:STATe] to be set OFF (if it is ON).
Name Type Range of Values Default Value
value numeric -254 to +252 volts
Range Maximum Value Resolution
0.0625
0.250
1.00
4.00
16.00
64.00
256.00
±0.062011719
±0.248046875
±0.992187500
±3.968750
±15.8750
±63.50
±254.00
0.000488281
0.001953125
0.00781250
0.031250
0.1250
0.500
2.0