Agilent Technologies E1564A Stereo Receiver User Manual


 
Digitizers Command Reference 59Chapter 3
CALibration:STORe
CALibration:STORe writes the calibration constants to non-volatile RAM
after calibration has been completed.
NOTE The FLASH and CAL CONSTANTS switches must be set to the “Write
Enable” positions before calibration constants are stored in RAM.
Executable when initiated: NO
Coupled Command: NO
Reset (*RST) Condition: None
CALibration:VALue
CALibration:VALue <voltage> specifies the voltage value actually applied
to the channel for calibration. This value informs the digitizer what voltage is
either being placed on the front panel input connector (CAL:SOURce
EXTernal) or the value being generated by the internal DAC (E1564A
4-Channel Digitizer only) and being output onto the calibration bus.
Parameters
Comments Source Maximum Voltages: The maximum voltage from an external source
used to calibrate the 16V range is 15.68V or 98% of full scale. The maximum
voltage attainable from the E1564A internal DAC is 15V.
Using the Internal DAC: The internal DAC on the E1564A can be used for the
calibration source when CAL:SOURce INTernal is specified. The output
level of this DAC is specified with CAL:DAC:VOLTage. The actual output
level must be measured with a voltmeter by the person doing the calibration.
That measured value is the value used for the <voltage> parameter of the
CAL:VALue command. The voltage can be measured across pins 5 (high)
and 9 (low) of the Calibration Bus Output (D-subminiature) calibration bus
connector.
Maximum Output Levels: The maximum output levels are limited to the levels
shown in the following table. These are the E1564A DAC voltages
recommended for calibrating each range. The values are approximately
98% of full scale (except for the 16V range which the internal E1564A’s
DAC has a maximum output of ±15V.
Name Type Range of Values Default Value
voltage numeric ±0.061256409 - ±15.6800 volts