148 Digitizers Error Messages Appendix C
-440 Query
UNTERMINATED
after indefinite
response
The *IDN? command must be the last query command within a
command string. Example: *IDN?;:SYST:VERS?
300 Not yet implemented NOT USED
1000 Illegal when initiated Many commands are not allowed to execute when the instrument is
busy taking a measurement - this error will occur if that is the case.
1001 Illegal while
calibrating
Many commands are not allowed to execute when the instrument is in
calibration mode. This error will occur when that is the case.
1002 Trigger ignored A valid trigger occurred, but was not expected at that time. Usually,
because a trigger has already been received for that measurement.
1003 Sample Trigger
ignored
This will occur if the instrument is taking a sample and a SAMPle:IMM
command is received during the previous sample period.
1004 Insufficient data for
query
This error will occur if you try to fetch readings, but have not initiated a
measurement, so no data is available.
1005 Invalid channel
number
This error usually is the result of trying to specify channels 3 or 4 for a
command sent to the E1563A (which only has two channels).
1006 Invalid channel range During data fetching, a channel range may be specified (e.g., (@1:4)).
A bad range could result by specifying (@1:5) since there is no
channel 5, or (@1:4) on a two-channel card like the E1563A, or a
descending range such as (@3:1).
1007 Error in CAL An error occurred while trying to perform the calibration command
specified.
1008 Data fetch timed out
waiting for trigger
Data fetch timed out waiting for trigger. It is allowable to ask for data
immediately after the INIT command. This error will occur if you ask for
data early, and the trigger that initiates the measurement has not been
received within the time period specified by the VISA timeout setting.
1009 Error reading bits,
viMoveIn16 failed
NOT USED
1010 Self test failed The abbreviated power on self test failed. A more thorough *TST self
test should be run for more specific information.
1011 VISA error An unknown error occurred in the VISA I/O library.
1012 Write to Flash ROM
failed
An unknown I/O error occurred while trying to write to flash ROM.
1013 Insufficient memory
for cal; try smaller
sample size
The controller did not have enough memory for the command to
complete, so try specifying a smaller sample size for the calibration
command that caused the error.
1014 Memory malloc failed;
insufficient memory
The controller did not have sufficient RAM for the command to
execute.
Number Title Description