Agilent Technologies E1564A Stereo Receiver User Manual


 
50 Digitizers Command Reference Chapter 3
CALCulate:LIMit:LOWer:DATA
CALCulate[<channel>]:LIMit:LOWer:DATA <value> | MIN | MAX sets the lower
limit value you want to test against. CALC<channel>:LIMit:FAIL? will return
a “1” following the measurement (and prior to the next measurement) if the
input signal fell below the specified lower limit value and if LIM:LOW:STATe
is ON. A “0” is returned if the limit was not exceeded.
Parameters
Comments Allowable Maximum Values: Allowable maximum values for the lower limit by
range and the associated resolution follow.
Executable when initiated: NO
Coupled Command: YES. Range changes will change the value. The
percent of full scale of the range will be kept constant. For example,
on the 4 volt range, with a 2V limit, a range change to 16V will set a new
limit of 8V.
Related Commands: [SENSe:]VOLTage[<channel>][:DC]:RANGe <range>
Reset (*RST) Condition: -254 volts
CALCulate:LIMit:LOWer:DATA?
CALCulate[<channel>]:LIMit:LOWer:DATA? [MIN | MAX] queries the lower
limit value set for the specified channel.
Name Type Range of Values Default Value
value numeric -254 to +252 volts
Range Maximum Value Resolution
0.0625
0.250
1.00
4.00
16.00
64.00
256.00
±0.061523438
±0.246093750
±0.984375000
±3.937500
±15.750
±63.00
±252.00
0.000488281
0.001953125
0.00781250
0.031250
0.1250
0.500
2.0