Incoming Inspection
CSA8000B & TDS8000B User Manual
1-23
CSA8000/TDS8000
SMA cable fromDC calibration
outputto 80E00 C3input
Figure 1- 7: Hookup for electrical functional tests
4. Set the DC CALIBRATOR OUTPUT:
a. Push the Vertical MENU front-panel button. This displays the Vert
Setup dialog box.
NOTE. When an optical module is installed, the optical setup dialog box displays
by default. Click the Basic button to display the basic dialog box.
b. Enter a level of 200 mV in the DC CAL box.
c. Push the Vertical MENU front-panel button again to dismiss the Vert
Setup dialog box.
5. Select the channel to test: Push the channel button for the channel you want
to test. The button lights and the channel display comes on. See Figure 1--8.
Channel
buttons
Figure 1- 8: Channel button location