Acquiring Waveforms
3-34
CSA8000B & TDS8000B User Manual
Overview Control el ements and resourcesTo acquire in FrameScan mode (cont.)
Set the bit rate 8. Set the horizontal scale so one acquisiti on record is
equal to one bit. Use one of t he two methods t hat follow:
H Automatic: If your signal to be scanned matches a
communications standard, select it from the Comm
Standard lis t. Choosing a standard sets the bit rate
and start bit; otherwise, if you know the bit rate, you
can set the bi t rate manually using the Bit Rate
box.
H Manual: Adjust the Scale control to a setting t hat
results in a display of both edges of the bit. For
example, setting 1/8 of a bit per division (0.125
bits/div) yields 1 bit in 8 divisions, which fits nicely
on screen.
Select acomm.
standard,or...
Set the bit r ate
manually
Set to1/8 bit
per division
Set the starting
horizontal posi-
tion
9. Set the initial horizontal position to the first bit you want
to acquire. Use one of t he two methods t hat follow:
H Automatic: Enter your desired start bit location, and
then check the Auto Position box to enable t he
instrument to set t he position as near as possi ble t o
match t he bit specified in the Start Bit box.
H Manual: Adjust the Position control to align t he
start of a bit to desired location in the frame.
Tip. The latter method is useful when you need to
manually align a bit or waveform to a m ask on the
display.
Or setmanually
First setthe
startbit,...
and then enable
Auto Position
Enable
FrameScan
10. In the dialog box, click to check the FrameScan
Enabled box. See right.
11. To restart the scan at the first bit at any time, click the
Reset button.
Click to
restartscan
Check to
startscan