Measuring Waveforms
3-94
CSA8000B & TDS8000B User Manual
Overview Related control elements and resourcesTo perform a com pensatio n (cont.)
Select the
scope of the
compensation
3. Wait until the Status for all items you want to
compensate changes from Warm Up to Comp Reqd or
Pass.
4. In the Select Action fields, select Compensate.
5. From the top pulldown li st, select the target to
compensate. Choose from:
H All to select the main instrument and all its
modules (defaul t selection).
H Mainframe to select only the main instrument.
H Module to select an individual module for
compensation.
If you have selected Module as the target, also choose
the channel to be compensated from the pulldown list of
channels.
Choose targetsto
compensate
Click to select
compensate
Enabled onlyif module
selected astarget
Run the
compensation
6. Click the Execute button to begin execution of t he
compensation.
Instructions t o disconnect inputs and i nstall dust
covers on opti cal module channels and 50
terminations on electrical module channels will
appear on screen. Be sure to f ollow stati c
precautions (see the user manual for your sampling
module) when following these instructions.
Note. Failing to install the 50 terminations can
yield erroneous compensation failures or results.
See Equipment Required on page 1--20.
The compensation may take several minut es t o
complete. Pass should appear as Status in the
dialog box when compensation completes.
If Fail appears as Status, rerun the compensation. If
Fail status cont inues af ter rerunning com pensation
and you have allowed warm up to occur, the module
or main instrument may need service.
Click to start
compensation