Tektronix TDS8000B Stereo System User Manual


 
Measuring Waveforms
CSA8000B & TDS8000B User Manual
3-95
Overview Related control elements and resourcesTo perform a com pensatio n (cont.)
Save the
compensation
7. In the Select Action fields, select Save.
8. Click the Execute button to save the new compensation
results. The new compensation results will be lost when
the instrum ent is powered down if they are not saved.
The Storage desti nation for the compensation results is
limited to the User area. The Factory settings cannot be
overwritten.
Recalling a
compensation
9. In the Select Action fields, select Recall.
10. In the Storage field, select the compensation to recall,
UserorFactory.
User recalls t he last saved user compensation. Factory
recalls the compensat ion established at the factory.
Note. Before proceeding, make sure you want to rewrite
the compensation you just saved.
11. Click the Execute button to recall the compensation
results.
Note. The Factory compensat ion should only be recalled
to bring the instrument to a known state. Recalling the
Factory compensation does not guarantee measurement
accuracy.
End of Procedure