Emerson 9739 Satellite Radio User Manual


 
6 Integrate the meter with the control
system
Topics covered in this chapter:
Configure the mA output
Configure the frequency output
Configure the discrete output
Configure the discrete input
Configure the mA input
Configure events
Configure digital communications
Set up polling for temperature
Set up polling for pressure
6.1 Configure the mA output
The mA output is used to report the configured process variable. The mA output
parameters control how the process variable is reported. Your transmitter has two mA
outputs.
The mA output parameters include:
mA Output Process Variable
Lower Range Value (LRV) and Upper Range Value (URV)
AO Cutoff
Added Damping
AO Fault Action and AO Fault Value
Important
Whenever you change an mA output parameter, verify all other mA output parameters before
returning the flowmeter to service. In some situations, the transmitter automatically loads a set of
stored values, and these values may not be appropriate for your application.
Integrate the meter with the control system
Configuration and Use Manual 83