- Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
- Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
• Before performing the calibration, record your current calibration parameters. If the
calibration fails, restore the known values.
Procedure
See Figure 9-4.
D3 or D3 and D4 density calibration using the Field CommunicatorFigure 9-4:
Enter density of D3 fluid
Density Calibration
Complete message
Calibration in Progress
message
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid Fill sensor with D4 fluid
D4 calibration
OK
Dens Pt 3 T-Series
Calibration method
executes
OK
Home
Enter density of D4 fluid
Density Calibration
Complete message
Calibration in Progress
message
OK
Dens Pt 4 T-Series
Calibration method
executes
OK
Home
Done
On-Line Menu >
Service Tools >
Maintenance >
Density Calibration
Service Tools >
Maintenance >
Density Calibration
Done
Measurement support
156 Micro Motion
®
9739 MVD Transmitters