Totalizer Start/Stop
•
Totalizer Reset
•
Acknowledge All Alarms
5.2.1 Enable or disable Totalizer Start/Stop from the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS STOP
ProLink II ProLink > Configuration > Display > Display Options > Display Start/Stop Totalizers
Field Communicator
Not available
Overview
You can control whether or not the operator is able to start and stop totalizers and
inventories from the display.
Restrictions
• You cannot start and stop totalizers individually from the display. All totalizers are started or
stopped together.
• You cannot start or stop inventories separately from totalizers. When a totalizer is started or
stopped, the associated inventory is also started or stopped.
• If the petroleum measurement application is installed on your computer, the operator must
enter the off-line password to perform this function, even if the off-line password is not
enabled.
Procedure
1. Ensure that at least one totalizer is configured as a display variable.
2. Enable or disable Totalizer Reset as desired.
Option
Description
Enabled
Operators can start and stop totalizers and inventories from the display,
if at least one totalizer is configured as a display variable.
Disabled (default)
Operators cannot start and stop totalizers and inventories from the dis-
play.
5.2.2 Enable or disable Totalizer Reset from the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS RESET
ProLink II ProLink > Configuration > Display > Display Options > Display Totalizer Reset
Field Communicator
Not available
Configure device options and preferences
70 Micro Motion
®
9739 MVD Transmitters