Emerson 9739 Satellite Radio User Manual


 
Status alarms and recommended actions (continued)Table 10-2:
Alarm code Description Cause Recommended actions
A005 Mass Flow Rate Over-
range
The measured flow has
exceeded the maxi-
mum flow rate of the
sensor (ΔT greater than
200 µs).
If other alarms are present, resolve those
alarm conditions first. If the current alarm per-
sists, continue with the recommended ac-
tions.
Check your process conditions against the val-
ues reported by the flowmeter.
Check for slug flow. See Section 10.25.
Check the drive gain and the pickoff voltage.
See Section 10.26 and Section 10.27.
Check for electrical shorts. See Section 10.28.
Check the integrity of the sensor tubes.
Contact Micro Motion.
A006 Characterization Re-
quired
Calibration factors have
not been entered and
the sensor type is incor-
rect.
Verify that all of the characterization parame-
ters match the data on the sensor tag.
Contact Micro Motion.
A008 Density Overrange The measured density
has exceeded
10 g/cm
3
.
If other alarms are present, resolve those
alarm conditions first. If the current alarm per-
sists, continue with the recommended ac-
tions.
Verify process conditions, checking especially
for air in the flow tubes, tubes not filled, for-
eign material in the tubes, or coating in the
tubes.
Check for slug flow. See Section 10.25.
If accompanied by an A003 alarm, check for
electrical shorts. See Section 10.28.
Verify that all of the characterization parame-
ters match the data on the sensor tag.
Check the drive gain and the pickoff voltage.
See Section 10.26 and Section 10.27.
Perform a density calibration.
Contact Micro Motion.
A009 Transmitter Initializing/
Warming Up
Transmitter is in power-
up mode.
Allow the meter to warm up.
Verify that the tubes are full of process fluid.
Check the wiring between the sensor and
transmitter. See Section 10.10.
A010 Calibration Failure Many possible causes,
such as too much flow
through the sensor dur-
ing a calibration proce-
dure.
If this alarm appears during zeroing, verify that
there is no flow through the sensor, then retry
the procedure.
Cycle power to the meter, then retry the pro-
cedure.
Troubleshooting
Configuration and Use Manual 161