Emerson 9739 Satellite Radio User Manual


 
Important
Whenever you change a discrete input parameter, verify all other discrete input parameters before
returning the flowmeter to service. In some situations, the transmitter automatically loads a set of
stored values, and these values may not be appropriate for your application.
6.4.1 Configure Discrete Input Action
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > DI > DI ACT
ProLink II ProLink > Configuration > Discrete Input > Assignment
Field Communicator Configure > Alert Setup > Discrete Events > Assign Discrete Action
Overview
Discrete Input Action controls the action or actions that the transmitter will perform when the
discrete input transitions from OFF to ON.
CAUTION!
Before assigning actions to an enhanced event or discrete input, check the status of the event
or the remote input device. If it is ON, all assigned actions will be performed when the new
configuration is implemented. If this is not acceptable, wait until an appropriate time to assign
actions to the event or discrete input.
Procedure
1. Select an action.
2. Select the discrete input that will perform the selected action.
3. Repeat until you have assigned all the actions to be performed by the discrete input.
Options for Discrete Input Action
Table 6-11:
Action Label
Display ProLink II ProLink III Field Communicator
Standard
None (default)
NONE None None None
Start sensor zero
START ZERO Start Sensor Zero Start Sensor Zero Perform auto zero
Start/stop all totaliz-
ers
START STOP Start/Stop All Totalization Start/Stop All Totalization Start/stop totals
Reset mass total
RESET MASS Reset Mass Total Reset Mass Total Reset mass total
Reset volume total
RESET VOL Reset Volume Total Reset Volume Total Reset volume total
Integrate the meter with the control system
104 Micro Motion
®
9739 MVD Transmitters