3.3 Enable access to the off-line menu of the
display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > OFFLN
ProLink II ProLink > Configuration > Display > Display Options > Display Offline Menu
Field Communicator
Not available
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must
enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from
another tool.
3.4 Disable write-protection on the transmitter
configuration
Display OFF-LINE MAINT > CONFG > LOCK
ProLink II ProLink > Configuration > Device > Enable Write Protection
Field Communicator
Not available
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it
before you can change any configuration parameters. By default, the transmitter is not
write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent
normal operational use. You can always disable write-protection, perform any required configuration
changes, then re-enable write-protection.
3.5
HART security
HART security may be enabled on your transmitter. To configure the transmitter using
HART protocol, you must disable HART security.
Introduction to configuration and commissioning
Configuration and Use Manual 19