Options for Volume Flow Measurement Unit for liquid applications (continued)Table 4-2:
Unit description
Label
Display ProLink II ProLink III Field Communica-
tor
Barrels per minute
(1)
BBL/MN barrels/min barrels/min bbl/min
Barrels per hour
(1)
BBL/H barrels/hr barrels/hr bbl/h
Barrels per day
(1)
BBL/D barrels/day barrels/day bbl/d
Beer barrels per second
(2)
BBBL/S Beer barrels/sec Beer barrels/sec bbbl/s
Beer barrels per minute
(2)
BBBL/MN Beer barrels/min Beer barrels/min bbbl/min
Beer barrels per hour
(2)
BBBL/H Beer barrels/hr Beer barrels/hr bbbl/h
Beer barrels per day
(2)
BBBL/D Beer barrels/day Beer barrels/day bbbl/d
Special unit
SPECL special special Spcl
Define a special measurement unit for volume flow
Display
Not available
ProLink II ProLink > Configuration > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Volume Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report
process data, totalizer data, and inventory data in a unit that is not available in the
transmitter. A special measurement unit is calculated from an existing measurement unit
using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to
select an existing special measurement unit, and to view process data using the special
measurement unit.
Procedure
1. Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
(2) Unit based on U.S. beer barrels (31 U.S. gallons).
Configure process measurement
Configuration and Use Manual 29