The GSV flow measurement parameters include:
Volume Flow Type
Standard Gas Density
Gas Standard Volume Flow Measurement Unit
Gas Standard Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. You
must choose one or the other.
4.3.1 Configure Volume Flow Type for gas applications
Display OFF-LINE MAINT > OFF-LINE CONFG > VOL > VOL TYPE GAS
ProLink II ProLink > Configuration > Flow > Vol Flow Type > Std Gas Volume
Field Communicator Configure > Manual Setup > Measurements > Gas Standard Volume > Volume Flow Type > GSV
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is
used.
Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2
Configure Standard Gas Density
Display
Not available
ProLink II ProLink > Configuration > Flow > Std Gas Density
Field Communicator Configure > Manual Setup > Measurements > Gas Standard Volume > Gas Density
Overview
The Standard Gas Density value is used to convert the measured flow data to the standard
reference values.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for
Standard Gas Density.
Procedure
Set Standard Gas Density to the standard reference density of the gas you are measuring.
Configure process measurement
32 Micro Motion
®
9739 MVD Transmitters