Status alarms and recommended actions (continued)Table 10-2:
Alarm code Description Cause Recommended actions
A113 mA Output 2 Saturated Check process conditions. Actual conditions
may be outside of the normal conditions for
which the output is configured.
Verify process conditions, checking especially
for air in the flow tubes, tubes not filled, for-
eign material in the tubes, or coating in the
tubes.
Verify that the measurement units are config-
ured correctly for your application.
Purge the flow tubes.
Check the settings of Upper Range Value and
Lower Range Value. See Section 10.17.
A114 mA Output 2 Fixed Check whether the output is in loop test
mode. If it is, unfix the output.
Exit mA output trim, if applicable.
Check whether the output has been set to a
constant value via digital communication.
A115 No External Input or
Polled Data
The HART polling con-
nection to an external
device has failed. No re-
sponse received from
polled device.
The mA input connec-
tion to an external de-
vice has failed. No re-
sponse received from
the external device.
Verify the external device operation.
Verify the wiring between the transmitter and
the external device.
Verify the HART polling configuration.
Verify the mA input configuration.
A116 Temperature Over-
range (Petroleum)
Check your process conditions against the val-
ues reported by the flowmeter.
Verify the configuration of the petroleum
measurement table type and temperature.
A117 Density Overrange (Pe-
troleum)
Check your process conditions against the val-
ues reported by the flowmeter.
Verify the configuration of the petroleum
measurement table type and density.
A118 Discrete Output 1 Fixed The discrete output has
been configured to
send a constant value.
Check whether the output is in loop test
mode. If it is, unfix the output.
A120 Curve Fit Failure (Con-
centration)
Verify the configuration of the concentration
measurement application.
A121 Extrapolation Alarm
(Concentration)
Check your process conditions against the val-
ues reported by the flowmeter.
Verify the configuration of the concentration
measurement application.
Troubleshooting
Configuration and Use Manual 165