4.8 Configure the concentration measurement
application
The concentration measurement application calculates concentration data from process
temperature and density. Micro Motion provides a set of concentration matrices that
provide the reference data for several standard industry applications and process fluids. If
desired, you can build a custom matrix for your process fluid, or purchase a custom matrix
from Micro Motion.
More information about the concentration measurement application is available in the
following manual: Micro Motion Enhanced Density Application: Theory, Configuration, and
Use.
Note
The concentration measurement application is also known as the enhanced density application.
4.8.1 Configure concentration measurement using ProLink II
This task guides you through loading and setting up a concentration matrix to use for
measurement. It does not cover building a concentration matrix.
Note
Concentration matrices can be made available on your transmitter either by loading an existing
matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter,
but only one can be used for measurement at any given time. See Micro Motion Enhanced Density
Application: Theory, Configuration, and Use for detailed information on building a matrix.
Prerequisites
Before you can configure concentration measurement:
• The concentration measurement application must be enabled on your transmitter.
• The concentration matrix you want to use must be available on your transmitter, or
it must be available as a file on your computer.
• You must know the derived variable that your matrix is designed for.
• You must know the density unit used by your matrix.
• You must know the temperature unit used by your matrix.
• The concentration measurement application must be unlocked.
Procedure
1. Choose ProLink > Configuration > Density and set Density Units to the density unit used by
your matrix.
2. Choose ProLink > Configuration > Temperature and set Temp Units to the temperature unit
used by your matrix.
3. Choose ProLink > Configuration > CM Setup.
Configure process measurement
Configuration and Use Manual 53