Tektronix TDS7254 Stereo System User Manual


 
Incoming Inspection
1-34
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
b. Click the Save button under Save settings to file in the control window.
This displays a familiar Windows dialog box for choosing a destination
directory naming the file.
c. In the Save Instrument Setup As dialog box, select the 3
1
/
2
Floppy (A:)
icon in the Save in: drop-down list to set the save destination to the
floppy disk.
d. Note the default file name, and then click the Save buttontosavethe
setup to the default file name.
9. Change the settings again: Set the horizontal SCALE to 200 s/div.
10. Verify the file system works:
a. Click the Recall Setups tab in the control window.
b. Click the Recall button under Recall settings from file in the control
window. This displays a familiar Windows dialog box for locating the
settings file that you want to recall.
c. In the Recall Instrument Setup dialog box, select the 3
1
/
2
Floppy (A:)
icon in the Look in: drop-down list.
d. Locate and double click in the dialog box on the setup file that you
previously stored.
e. Verify that the instrument retrieved the saved setup from the disk. Do
this by noticing the horizontal SCALE is again 1 ms and the waveform
shows ten cycles just as it did when you saved the setup.
11. Remove the test hookup:
a. Disconnect the equipment from the channel input and the probe
compensation output.
b. Remove the floppy disk from the floppy disk drive.