Tektronix TDS7254 Stereo System User Manual


 
Appendix A: Specifications
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
A-19
Table A- 3: Horizontal and acquisition system specifications (Cont.)
Characteristic Description
Maximum record length, sample mode Depends on the number of active channels and the record length options inst alled.
Maximum record length is less in serial trigger mode
CSA7404, CSA7154,
TDS7404, TDS7254, TDS7154,
TDS7104, & TDS7054
Depends on the number of active channels and the record length options installed.
Maximum record length is less in serial trigger mode
Standard 500,000 points (3 or 4 channels)
1
0
0
0
0
0
0
p
o
i
n
t
s
(
1
o
r
2
c
h
a
n
n
e
l
s
o
n
l
y
)
1,000,000 po
i
nts
(
1or2
c
h
a
nne
l
son
l
y
)
2,000,000 points (1 channel only)
Option 2M instal led 2,000,000 points (3 or 4 channels)
4,000,000 points (1 or 2 channel s only)
8,000,000 points (1 channel only)
Option 3M instal led 4,000,000 points (3 or 4 channels)
8,000,000 points (1 or 2 channel s only)
16,000,000 points (1 channel only)
Option 4M instal led
CSA7404, CSA7154,
TDS7404, TDS7254, & TDS7154
8,000,000 points (3 or 4 channel s)
16,000,000 points (1 or 2 channel s only)
32,000,000 points (1 channel only)
TDS6604, & TDS6404 Maximum record length supported by acquisition hardware, at any sample rate
Mode Channels Record length
Sample 1 channel only 250,000
2 channels only:
channel 1 with 3,
channel 1 with 4,
channel 2 with 3,
or channel 2 with 4
250,000
2 channels only:
channel 1 with 2,
or channel 3 with 4
125,000
3 or 4 channels 125,000
Hi Res Any 12,500