Acquiring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3-57
FastFrame is not compatible with these fe atur es or modes:
H Equivalent Time
H Histograms
H Fast Acquisitions
H Average
H Envelope
H Waveform Database
Consider the following operating characteristics when using FastFrame:
H You can push RUN/STOP to terminate a FastF rame sequence. If any frames
were acquired, they are displayed. If no frames were acquired, the previous
FastFrame waveform is displayed.
H Because FastFrame introduces additional processing time into the operation
cycle of acquire, process, and display, its best to use Single Sequence
Acquisition (see Acquire menu, Stop After menu). With Single Sequence
selected, you will see the current acquisition sequence; otherwise, the display
lags the current sequence by one sequence. You can also see the current
sequence by pushing the RUN/STOP button to stop the acquisition.
Using FastFrame
Acquisitions