Pressure Compensation and Temperature Compensation
There are two ways to configure external temperature compensation:
If the operating temperature is a known static value, you can configure that value in the
transmitter.
If the operating temperature varies significantly, you must write a temperature value to the
transmitter at appropriate intervals, using an appropriate output assembly. See Section 9.4.
Note: Ensure that your temperature value is accurate, or that your temperature measurement device is
accurate and reliable.
To enable and configure external temperature compensation:
With ProLink II, see Figure 9-3.
With a DeviceNet tool, see Figure 9-4.
Figure 9-3 Configuring external temperature compensation with ProLink II
Configure
Enable Use External
Temperature
Enable
Apply
Enter Temperature units
(1)
Apply
View Menu >
Preferences
ProLink >
Configuration >
Temperature
Set up output
assembly
(2)
Enter External
Temperature
Yes
Done
No
Use static
temp value?
Apply
(1) Temperature measurement unit must be
configured to match temperature unit
used by external device. See Section 6.3.
(2) See Section 9.4.