108 Micro Motion
®
Model 2400S Transmitters for DeviceNet
Measurement Performance
Figure 10-11 D3 or D3 and D4 density calibration – DeviceNet tool
Enter density of
D3 fluid
Monitor status
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with
D3 fluid
Fill sensor with
D4 fluid
D4 calibration
Done
Object: Calibration Object (0x65)
Instance: 1
Service: 0x4F
Start D3
calibration
Object: Calibration Object (0x65)
Instance: 1
Attribute ID: 15
Data type: REAL
Service: Set
Yes
Object: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 16
Bit: 7
Service: Get
Bit 7 0ff?No
Enter density of
D4 fluid
Monitor status
Object: Calibration Object (0x65)
Instance: 1
Service: 0x50
Start D4
calibration
Object: Calibration Object (0x65)
Instance: 1
Attribute ID: 16
Data type: REAL
Service: Set
Object: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 16
Bit: 6
Service: Get
Bit 6 0ff?
Yes
No
Done