Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool
8.9 Configuring the display
If your transmitter has a display, you can configure a variety of parameters that control the display
functionality.
8.9.1 Update period
The Update Period (or Display Rate) parameter controls how often the display is refreshed with
current data. The default is
200 milliseconds; the range is 100 milliseconds to 10,000 milliseconds
(10 seconds).
To configure Update Period:
Using ProLink II, see Figure B-3.
Using the display menus, see Figure B-6.
Using a DeviceNet tool, see Table C-9.
8.9.2 Language
The display can be configured to use any of the following languages for data and menus:
English
•French
•German
Spanish
To set the display language:
Using ProLink II, see Figure B-3.
Using the display menus, see Figure B-6.
Using a DeviceNet tool, see Table C-9.
8.9.3 Enabling and disabling display functions
Table 8-9 lists the display functions and describes their behavior when enabled (shown) or disabled
(hidden).
A116 API: Temperature Outside Standard Range Info Yes No
A117 API: Density Outside Standard Range Info Yes No
A120 ED: Unable to Fit Curve Data Info No No
A121 ED: Extrapolation Alarm Info Yes No
A131 Meter Verification/Outputs at Last Value Info Yes No
A132 Simulation Mode Active Info Yes No
A133 PIC UI EEPROM Error Info Yes No
(1) Can be set to either Informational or Ignore, but cannot be set to Fault.
Table 8-8 Status alarms and severity levels continued
Alarm code ProLink II message
Default
severity Configurable?
Affected by
fault timeout?