Emerson Process Management 2400S Satellite Radio User Manual


 
Configuration and Use Manual 95
Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
Table 10-1 DeviceNet interface for meter verification
Step number Step description Interface
1 Set output state Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 58
Data type: USINT
Value:
• 0: Last measured value (default)
• 1: Fault
Service: Set
2 Set uncertainty limit Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 59
Data type: REAL
Range: 0.0025 to 0.05
Default: 0.04
Service: Set
3 Start/abort procedure Class: Diagnostics Object (0x66)
Instance: 1
• 0: Abort
•1: Start
Service: 0x50
4 Check current algorithm state Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 54
Data type: USINT
Service: Get
5 Read percent complete Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 57
Data type: USINT
Service: Get
6 Check algorithm abort state Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 56
Data type: USINT
Service: Get
7 Check inlet stiffness Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 61
Data type: USINT
• 0: Within uncertainty limit
• 1: Outside uncertainty limit
Service: Get
8 Check outlet stiffness Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 62
Data type: USINT
• 0: Within uncertainty limit
• 1: Outside uncertainty limit
Service: Get
9 Read abort code Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 55
Data type: USINT
Codes: See Table C-7
Service: Get