Model 2400S Transmitters for DeviceNet
Device Profile
55 Meter
verification abort
code
USINT Get V The reason the meter
verification routine
aborted
•0 = No error
• 1 = Manual abort
• 2 = Watchdog timeout
• 3 = Frequency drift
• 4 = High peak drive
voltage
• 5 = High drive current
standard deviation
• 6 = High drive current
mean
• 7 = Drive loop
reported error
• 8 = High Delta T
standard deviation
• 9 = High Delta T
value
• 10 = State running
• 11 = Verification
complete
• 12 = Wrong
verification enable
• 13 = No factory air
verification
• 14 = No factory water
verification
• 15 = Parameters not
set
56 Meter
verification
algorithm state
at abort
USINT Get V The state of the meter
verification routine when it
aborted
1–18
57 Meter
verification
percent
complete
USINT Get V The progress of the meter
verification routine
%
58 Meter
verification
outputs state
USINT Set NV The state of the outputs
when the meter
verification routine is
running
• 0 = Last value
•1 = Fault
59 Meter
verification
stiffness limit
REAL Set NV The setpoint of the
stiffness limit. Represents
percentage.
Unitless
60 Meter
verification
validation
counter
UINT Get NV Indicates the number of
times the meter
verification routine has
successfully completed
61 Meter
verification inlet
stiffness out of
limits
USINT Get V Is the inlet stiffness out of
limits?
•0 = No
•1 = Yes
62 Meter
verification
outlet stiffness
out of limits
USINT Get V Is the outlet stiffness out
of limits?
•0 = No
•1 = Yes
Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Attrib
ID Name Data type Service Mem Description Comments