Emerson Process Management 2400S Satellite Radio User Manual


 
40 Micro Motion
®
Model 2400S Transmitters for DeviceNet
Using the Transmitter
16
(1)(3)
Status
• API temperature-corrected density
API temperature-corrected volume flow
• API temperature-corrected volume inventory
• API average temperature-corrected density
• API average temperature
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Petroleum
measurement
application
17
(1)(4)
Status
Mass flow
• Volume flow
• Temperature
• Enhanced density reference density
• Enhanced density specific gravity
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
18
(1)(4)
Status
Mass flow
• Volume flow
• Temperature
• Density
• Enhanced density concentration
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
19
(1)(4)
Status
Mass flow
• Volume flow
• Temperature
• Density
• Enhanced density Baume
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
20
(4)
• Status
• Temperature
• Density
• Enhanced density net mass flow
• Enhanced density net mass total
• Enhanced density net mass inventory
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
21
(4)
• Status
• Temperature
• Density
• Enhanced density net volume flow
• Enhanced density net volume total
• Enhanced density net volume inventory
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
22
(4)
• Status
•Mass Flow
• Temperature
• Density
• Enhanced density reference density
• Enhanced density net mass flow
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
23
(1)(4)
• Status
• Volume flow
• Temperature
• Density
• Enhanced density reference density
• Enhanced density net volume flow
21 • BOOL
• REAL
• REAL
• REAL
• REAL
• REAL
Enhanced
density
application
Table 7-2 Summary of input assemblies continued
Instance ID Data description Size (bytes) Data type Description