Configuration and Use Manual 103
Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
Figure 10-7 DeviceNet tool – Flowmeter zero procedure
10.6 Performing density calibration
Density calibration includes the following calibration points:
• All sensors:
- D1 calibration (low-density)
- D2 calibration (high-density)
• T-Series sensors only:
- D3 calibration (optional)
- D4 calibration (optional)
For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density
measurement. If you choose to perform the D3 and D4 calibration:
• Do not perform the D1 or D2 calibration.
• Perform D3 calibration if you have one calibrated fluid.
• Perform both D3 and D4 calibrations if you have two calibrated fluids (other than air and
water).
Modify zero time if
required
Perform zero
Class: Calibration Object (0x65)
Instance: 1
Attribute ID: 3
Data type: UINT
Units: seconds
Service: Set
Class: Calibration Object (0x65)
Instance: 1
Service: 0x4B (perform or abort zero)
Value:
· 0: Abort zero calibration
· 1: Start zero calibration
Check outcome
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID 12, Bit 0x0020
Value:
· 0: Zero succeeded
· 1: Zero failed
Data type: USINT
Check zero values
Class: Calibration Object (0x65)
Instance: 1
Attribute ID 4: Zero standard deviation
Attribute ID 5: Zero offset
Data type: REAL
Units: milliseconds
Check status
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID 16, Bit 0x8000
Value:
· 0: Zero complete
· 1: Zero in progress
Data type: USINT