Configuration and Use Manual 103
Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
Figure 10-7 DeviceNet tool – Flowmeter zero procedure
10.6 Performing density calibration
Density calibration includes the following calibration points:
All sensors:
- D1 calibration (low-density)
- D2 calibration (high-density)
T-Series sensors only:
- D3 calibration (optional)
- D4 calibration (optional)
For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density
measurement. If you choose to perform the D3 and D4 calibration:
Do not perform the D1 or D2 calibration.
Perform D3 calibration if you have one calibrated fluid.
Perform both D3 and D4 calibrations if you have two calibrated fluids (other than air and
water).
Modify zero time if
required
Perform zero
Class: Calibration Object (0x65)
Instance: 1
Attribute ID: 3
Data type: UINT
Units: seconds
Service: Set
Class: Calibration Object (0x65)
Instance: 1
Service: 0x4B (perform or abort zero)
Value:
· 0: Abort zero calibration
· 1: Start zero calibration
Check outcome
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID 12, Bit 0x0020
Value:
· 0: Zero succeeded
· 1: Zero failed
Data type: USINT
Check zero values
Class: Calibration Object (0x65)
Instance: 1
Attribute ID 4: Zero standard deviation
Attribute ID 5: Zero offset
Data type: REAL
Units: milliseconds
Check status
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID 16, Bit 0x8000
Value:
· 0: Zero complete
· 1: Zero in progress
Data type: USINT