Emerson Process Management 2400S Satellite Radio User Manual


 
Configuration and Use Manual 155
Device Profile
Device Profile IndexDisplay CodesMenus
41 Alarm 8 UINT Get V A collection of status bits 0x0001 = Not used
• 0x0002 = Not used
• 0x0004 = Not used
• 0x0008 = Not used
• 0x0010 = Not used
• 0x0020 = Not used
• 0x0040 = Not used
• 0x0080 = Not used
• 0x0100 = Not used
• 0x0200 = Not used
• 0x0400 = Not used
• 0x0800 = Not used
• 0x1000 = Not used
• 0x2000 = Not used
• 0x4000 = Not used
• 0x8000 = Not used
42 Alarm status USINT Set V The status of the alarm
selected in Attribute 18.
Write 0x00 to
acknowledge the alarm
selected in Attribute 18.
• 0x00 = Acked
/Cleared
• 0x01 = Acked/Active
•0x10 = Not
Acked/Cleared
•0x11 = Not
Acked/Active
43 Alarm count UINT Get V The number of inactive-to-
active transitions of the
alarm selected in
Attribute 18
44 Alarm last
posted
Unsigned 32 Get V The number of seconds
since the last reset that
the alarm selected in
Attribute 18 was posted
Seconds since last
reset
45 Alarm last
cleared
Unsigned 32 Get V The number of seconds
since the last reset that
the alarm selected in
Attribute 18 was cleared
Seconds since last
reset
46 Alarm history
index
USINT Set V The entry in the alarm
history log
Range: 0–49
47 Alarm history
alarm number
USINT Get V The alarm number that
corresponds to the alarm
history entry selected in
Attribute 45
1 = A001, 2 = A002,
etc.
48 Alarm history
alarm status
changed
USINT Get V The alarm status change
that corresponds to the
alarm history entry
selected in Attribute 45
•1 = Posted
• 2 = Cleared
49 Alarm history
alarm status
changed
timestamp
Unsigned 32 Get V The timestamp of the
alarm status change that
corresponds to the alarm
history entry selected in
Attribute 45
Seconds since last
reset
54 Meter
verification
algorithm state
USINT Get V The current state of the
meter verification routine
1–18
Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Attrib
ID Name Data type Service Mem Description Comments