106 Micro Motion
®
Model 2400S Transmitters for DeviceNet
Measurement Performance
Figure 10-9 D1 and D2 density calibration – DeviceNet tool
Enter density of
D1 fluid
Monitor status
D1 calibration
Close shutoff valve
downstream from sensor
Fill sensor with
D1 fluid
Fill sensor with
D2 fluid
D2 calibration
Done
Object: Calibration object (0x65)
Instance: 1
Service: 0x4C
Start D1
calibration
Object: Calibration object (0x65)
Instance: 1
Attribute ID: 12
Data type: REAL
Service: Set
Yes
Object: Diagnostics object (0x66)
Instance: 1
Attribute ID: 16
Bit: 14
Service: Get
Bit 14 0ff?No
Enter density of
D2 fluid
Monitor status
Object: Calibration object (0x65)
Instance: 1
Service: 0x4D
Start D2
calibration
Object: Calibration object (0x65)
Instance: 1
Attribute ID: 13
Data type: REAL
Service: Set
Object: Diagnostics object (0x66)
Instance: 1
Attribute ID: 16
Bit: 13
Service: Get
Bit 13 0ff?
Yes
No