Emerson Process Management 2400S Satellite Radio User Manual


 
Configuration and Use Manual 63
Optional Configuration
Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool Required Configuration Optional ConfigurationUsing the TransmitterUsing a DeviceNet Tool
Reset all totals RESET ALL 8 Resets the value of all totalizers to 0
Start/stop all totalization START STOP 9 If totalizers are running, stops all totalizers
If totalizers are not running, starts all totalizers
Increment current ED
curve
INCR CURVE 18 Changes the active enhanced density curve from 1 to
2, from 2 to 3, etc.
(4)
(1) Displayed only if Volume Flow Type = Liquid.
(2) Displayed only if Volume Flow Type = Gas.
(3) Available only if the petroleum measurement application is installed.
(4) Available only if the enhanced density application is installed.
Example
Define Event 1 to be active when the mass flow rate in forward or
backward direction is less than 2 lb/min. Additionally, if this occurs, all
totalizers should be stopped.
Using ProLink II:
1. Specify lb/min as the mass flow unit. See Section 6.3.1.
2. Configure the Flow Direction parameter for bidirectional flow. See
Section 8.5.
3. Select Event 1.
4. Configure:
Event Type = Low
Process Variable (PV) = Mass Flow Rate
Low Setpoint (A) = 2
5. In the Discrete Input panel, open the dropdown list for Start/Stop
All Totalization and select Discrete Event 1.
Using a DeviceNet tool:
1. Specify lb/min as the mass flow unit. See Section 6.3.1.
2. Configure the Flow Direction parameter for bidirectional flow. See
Section 8.5.
3. In the Diagnostics Object (0x66), Instance 1, set the following
attributes:
Discrete event index (Attribute 6) = 0
Discrete event action (Attribute 7) = 1
Discrete event process variable (Attribute 10) = 0
Discrete event setpoint A (Attribute 8) = 2
Discrete event action code (Attribute 84) = 9
Discrete event assignment (Attribute 85) = 57
Table 8-6 Event actions continued
ProLink II label Display label
DeviceNet
code Description