Table C-2 Analog Input Point Object (0x0A) – Instance 2 (liquid volume flow)
Attrib
ID Name Data type Service Mem Description Comments
3 Value REAL Get V Current value of liquid
volume flow process
variable
Based on Attribute 8
4 Status BOOL Get V Point status 0 = Good
• 1 = Alarm state
8 Value data type USINT Get V Data type used to report
volume flow process
variable
1 (REAL)
100 Process total REAL Get
Reset
(1)
(1) Service code 0x4B.
V Current value of liquid
volume total
101 Inventory total REAL Get
Reset
(2)
(2) Service code 0x4C.
V Current value of liquid
volume inventory
102 Value
engineering
units
UINT Set NV Liquid volume flow
measurement unit
See Table 6-3 for unit
codes.
103 Total
engineering
units
UINT Get V Liquid volume total and
inventory units
Transmitter
automatically
determines this based
on Attribute 102.
See Table C-13 for unit
codes.
105 Cutoff REAL Set NV Value below which liquid
volume flow will be
reported as 0
106 Meter factor REAL Set NV A multiplier to the
calculated liquid volume
flow
108 Reset volume
total
USINT Set V Resets the volume total 1 = Reset
109 Reset volume
inventory
USINT Set V Resets the volume
inventory
• 1 = Reset
Table C-3 Analog Input Point Object (0x0A) – Instance 3 (density)
Attrib
ID Name Data type Service Mem Description Comments
3 Value REAL Get V Current value of density
process variable
Based on Attribute 8
4 Status BOOL Get V Point status 0 = Good
• 1 = Alarm state
8 Value data type USINT Get V Data type used to report
density process variable
1 (REAL)
102 Value
engineering
units
UINT Set NV Density measurement unit See Table 6-5 for unit
codes.