Emerson MMI-20019043 Satellite Radio User Manual


 
3. To require a password for access to the maintenance section of the off-line menu
and the Smart Meter Verification menu, enable or disable Off-Line Password.
Option Description
Enabled
Operator is prompted for the off-line password at entry to the Smart Meter
Verification menu (if applicable) or entry to the maintenance section of the
off-line menu.
Disabled (de-
fault)
No password is required for entry to the Smart Meter Verification menu (if
applicable) or entry to the maintenance section of the off-line menu.
4. To require a password to access the alarm menu, enable or disable Alarm Password.
Option Description
Enabled
Operator is prompted for the off-line password at entry to the alarm
menu.
Disabled (default)
No password is required for entry to the alarm menu.
If both Off-Line Password and Alarm Password are enabled, the operator is prompted for
the off-line password to access the off-line menu, but is not prompted thereafter.
5.
(Optional) Set Off-Line Password to the desired value.
The same value is used for both the off-line password and the alarm password. The
default value is 1234. The range is 0000 to 9999.
Tip
Record your password for future reference.
5.4 Configure response time parameters
You can configure the rate at which process data is polled and process variables are
calculated.
Response time parameters include:
Update Rate
Calculation Speed (Response Time)
Configure device options and preferences
86 Micro Motion
®
Model 2700 Transmitters with Analog Outputs