API reference tables, associated process fluids, and associated calculation values (continued)Table 4-13:
Table
name Process fluid CTL source data Reference temperature Density unit
53B Generalized products Observed density and
observed temperature
15 °C (configurable) Base density
Range: 653 to
1075 kg/m
3
53D Lubricating oils Observed density and
observed temperature
15 °C (configurable) Base density
Range: 825 to
1164 kg/m
3
54C Liquids with a constant
density base or known
thermal expansion coef-
ficient
User-supplied reference
density (or thermal ex-
pansion coefficient) and
observed temperature
15 °C (configurable) Base density in kg/m
3
4.8 Configure the concentration measurement
application
The concentration measurement application calculates concentration data from process
temperature and density. Micro Motion provides a set of concentration matrices that
provide the reference data for several standard industry applications and process fluids. If
desired, you can build a custom matrix for your process fluid, or purchase a custom matrix
from Micro Motion.
More information about the concentration measurement application is available in the
following manual: Micro Motion Enhanced Density Application: Theory, Configuration, and
Use.
Note
The concentration measurement application is also known as the enhanced density application.
4.8.1 Configure concentration measurement using ProLink II
This task guides you through loading and setting up a concentration matrix to use for
measurement. It does not cover building a concentration matrix.
Note
Concentration matrices can be made available on your transmitter either by loading an existing
matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter,
but only one can be used for measurement at any given time. See Micro Motion Enhanced Density
Application: Theory, Configuration, and Use for detailed information on building a matrix.
Prerequisites
Before you can configure concentration measurement:
Configure process measurement
58 Micro Motion
®