Emerson MMI-20019043 Satellite Radio User Manual


 
For D3 density calibration, the D3 fluid must meet the following requirements:
- Minimum density of 0.6 g/cm
3
- Minimum difference of 0.1 g/cm
3
between the density of the D3 fluid and the
density of water. The density of the D3 fluid may be either greater or less than
the density of water.
For D4 density calibration, the D4 fluid must meet the following requirements:
- Minimum density of 0.6 g/cm
3
- Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
- Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
Before performing the calibration, record your current calibration parameters. If the
calibration fails, restore the known values.
Procedure
See Figure 11-12.
Measurement support
202 Micro Motion
®
Model 2700 Transmitters with Analog Outputs