Temperature calibration using ProLink IIIFigure 11-14:
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Start Calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Done
Device Tools >
Calibration >
Temperature Calibration >
Temperature Calibration - Offset
Start Calibration
Device Tools >
Calibration >
Temperature Calibration >
Temperature Calibration - Slope
Measurement support
Configuration and Use Manual 205