Tag on older straight-tube sensor (T-Series)Figure 2-3:
Tag on newer straight-tube sensor (T-Series)Figure 2-4:
Density calibration parameters (D1, D2, K1, K2, FD, DT, TC)
If your sensor tag does not show a D1 or D2 value:
• For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air. If
you cannot find a Dens A or D1 value, enter 0.001 g/cm
3
.
• For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water.
If you cannot find a Dens B or D2 value, enter 0.998 g/cm
3
.
If your sensor tag does not show a K1 or K2 value:
• For K1, enter the first 5 digits of the density calibration factor. In the sample tag, this
value is shown as 12500 (see Figure 2-1).
• For K2, enter the second 5 digits of the density calibration factor. In the sample tag,
this value is shown as 14286 (see Figure 2-1).
If your sensor does not show an FD value, contact Micro Motion customer service.
Quick start
12 Micro Motion
®
Model 2700 Transmitters with Analog Outputs