• Current flowmeter identification data
• Current flow and density configuration parameters
• Current zero values
• Current process values for mass flow rate, volume flow rate, density, temperature,
and external pressure
• Customer and test descriptions (if entered by the user)
If you use ProLink II or ProLink III to run a test, a test result chart and a test report are
displayed at the completion of the test. On-screen directions are provided to manipulate
the test data or export the data to a CSV file for offline analysis.
View test result data using the display
1. If you have just run a test, results are displayed automatically at the end of the test.
2. If you want to view results from previous tests:
a. Navigate to the Smart Meter Verification menu.
Smart Meter Verification – Top-level menuFigure 11-3:
Scroll and Select simultaneously
for 4 seconds
ENTER METER VERFY
Scroll
RUN VERFY RESULTS READ SCHEDULE VERFY
Select
EXIT
Scroll Scroll Scroll
Scroll Select
Select Select Select
b. Scroll to Results Read and press Select.
The runcount of the most recent test is displayed.
c.
To view data for this test, press Select, then press Scroll to scroll through test data.
d. To select a different test, press Scroll, then press Select when the transmitter
displays Results More?. When the desired test appears, as identified by runcount,
press Select.