Status alarms and Status Alarm SeverityTable 5-2:
Alarm code Status message Default severity Notes Configurable?
A001 EEPROM Error (Core Pro-
cessor)
Fault
No
A002 RAM Error (Core Processor)
Fault
No
A003 No Sensor Response
Fault
Yes
A004 Temperature Overrange
Fault
No
A005 Mass Flow Rate Overrange
Fault
Yes
A006 Characterization Required
Fault
Yes
A008 Density Overrange
Fault
Yes
A009 Transmitter Initializing/
Warming Up
Fault
Yes
A010 Calibration Failure
Fault
No
A011 Zero Calibration Failed:
Low
Fault
Yes
A012 Zero Calibration Failed:
High
Fault
Yes
A013 Zero Calibration Failed:
Unstable
Fault
Yes
A014 Transmitter Failure
Fault
No
A016 Sensor RTD Failure
Fault
Yes
A017 T-Series RTD Failure
Fault
Yes
A018 EEPROM Error (Transmit-
ter)
Fault
No
A019 RAM Error (Transmitter)
Fault
No
A020 No Flow Cal Value
Fault
Yes
A021 Incorrect Sensor Type (K1)
Fault
No
A022 Configuration Database
Corrupt (Core Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A023 Internal Totals Corrupt
(Core Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A024 Program Corrupt (Core
Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A025 Boot Sector Fault (Core
Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A026 Sensor/Transmitter Com-
munications Failure
Fault
No
A027 Security Breach
Fault
No
A028 Core Processor Write Fail-
ure
Fault
No
Configure device options and preferences
92 Micro Motion
®
Model 2700 Transmitters with Analog Outputs