Emerson MMI-20019043 Satellite Radio User Manual


 
Options for Enhanced Event Action
Options for Enhanced Event ActionTable 6-12:
Action Label
Display ProLink II ProLink III Field Communicator
Standard
None (default)
NONE None None None
Start sensor zero
START ZERO Start Sensor Zero Start Sensor Zero Perform auto zero
Start/stop all totaliz-
ers
START STOP Start/Stop All Totalization Start/Stop All Totalization Start/stop totals
Reset mass total
RESET MASS Reset Mass Total Reset Mass Total Reset mass total
Reset volume total
RESET VOL Reset Volume Total Reset Volume Total Reset volume total
Reset gas standard
volume total
RESET GSVT Reset Gas Std Volume
Total
Reset Gas Std Volume
Total
Reset gas standard vol-
ume total
Reset all totals
RESET ALL Reset All Totals Reset All Totals Reset totals
Petroleum measurement
Reset temperature-
corrected volume to-
tal
TCVOL Reset API Ref Vol Total Reset Volume Total at
Reference Temperature
Reset corrected volume
total
Concentration measurement
Reset CM reference
volume total
RESET STD V Reset CM Ref Vol Total Reset Volume Total at
Reference Temperature
Not available
Reset CM net mass to-
tal
RESET NET M Reset CM Net Mass Total Reset Net Mass Total
Not available
Reset CM net volume
total
RESET NET V Reset CM Net Vol Total Reset Net Volume Total
Not available
Increment CM matrix
INCr CURVE Increment Current CM
Curve
Increment Concentration
Matrix
Not available
Meter verification
Start meter verifica-
tion test
START VERFY Start Meter Verification Start Meter Verification
Not available
CAUTION!
Before assigning actions to an enhanced event or discrete input, check the status of the event
or the remote input device. If it is ON, all assigned actions will be performed when the new
configuration is implemented. If this is not acceptable, wait until an appropriate time to assign
actions to the event or discrete input.
Integrate the meter with the control system
Configuration and Use Manual 123