Procedure
Set Volume Flow Type to Liquid.
4.2.2 Configure Volume Flow Measurement Unit for liquid
applications
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL
ProLink II ProLink > Configuration > Flow > Vol Flow Units
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Overview
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the
volume flow rate. The unit used for the volume total and volume inventory is based on this
unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to
Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid
applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement
Unit, plus one user-defined measurement unit. Different communications tools may use
different labels for the units.
Options for Volume Flow Measurement Unit for liquid applicationsTable 4-3:
Unit description
Label
Display ProLink II ProLink III Field Communica-
tor
Cubic feet per second
CUFT/S ft3/sec ft3/sec Cuft/s
Cubic feet per minute
CUF/MN ft3/min ft3/min Cuft/min
Configure process measurement
30 Micro Motion
®
Model 2700 Transmitters with Analog Outputs