FVZ is a diagnostic variable that monitors the zero value over a period of 3 minutes.
It must be read during meter commissioning to comply with MID requirements for
Weights & Measures applications in Measuring Instrument Directive (MID) 2004/22/
EC. Refer to your Standard Operating Procedures documentation for detailed
instructions.
9. Enable software security:
a. Choose Plug-ins > Enable/Disable Custody Transfer.
b. Click Enable Custody Transfer.
Important
Depending on local requirements, this step may need to be witnessed by a certified Weights
& Measures inspector.
After this step, the transmitter is in a condition of metrological security
(measurement security). The transmitter performs the following actions:
Securing all measurement parameters. You can read the current configurations
but you cannot change them.
Clearing Status Alarm A027: Security Breach.
10.
Install the physical seal.
Important
In most installations, the physical seal is a wire seal that must be installed by a certified
Weights & Measures inspector. The seal is provided by the inspector. The physical seal is
inserted through the locking clamps on the transmitter (if available on your transmitter).
Example of locking clamp and seal on transmitterFigure 8-1:
Set up the Weights & Measures application
142 Micro Motion
®
Model 2700 Transmitters with Analog Outputs