• If you plan to configure an output to report a concentration measurement process
variable, ensure that the concentration measurement application is configured so
that the desired variable is available.
• If you are using the HART variables, be aware that changing the configuration of mA
Output Process Variable will change the configuration of the HART Primary Variable
(PV).
• If you have configured Display Variable 1 to track mA Output Process Variable, be aware
that changing the configuration of mA Output Process Variable will change the contents
of Display Variable 1.
Procedure
Set mA Output Process Variable as desired.
The default setting is Mass Flow Rate.
Options for mA Output Process Variable
The transmitter provides a basic set of options for mA Output Process Variable, plus several
application-specific options. Different communications tools may use different labels for
the options.
Options for mA Output Process VariableTable 6-1:
Process variable Label
Display ProLink II ProLink III Field Communicator
Standard
Mass flow rate
MFLOW Mass Flow Rate Mass Flow Rate Mass flo
Volume flow rate
VFLOW Volume Flow Rate Volume Flow Rate Vol flo
Gas standard volume
flow rate
GSV F Gas Std Vol Flow Rate Gas Standard Volume
Flow Rate
Gas vol flo
Temperature
TEMP Temperature Temperature Temp
Density
DENS Density Density Dens
External pressure
EXT P External Pressure External Pressure External pres
External temperature
EXT T External Temperature External Temperature External temp
Drive gain
DGAIN Drive Gain Drive Gain Driv signl
Petroleum measurement
Temperature-correc-
ted density
TCDEN API: Temp Corrected
Density
Density at Reference
Temperature
TC Dens
Temperature-correc-
ted (standard) volume
flow rate
TCVOL API: Temp Corrected Vol-
ume Flow
Volume Flow Rate at Ref-
erence Temperature
TC Vol
Average corrected
density
AVE D API: Avg Density Average Density TC Avg Dens
Average temperature
AVE T API: Avg Temperature Average Temperature TC Avg Temp
Integrate the meter with the control system