Emerson MMI-20019043 Satellite Radio User Manual


 
4.5.1 Configure Density Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS
ProLink II ProLink > Configuration > Density > Density Units
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit specifies the units of measure that will be displayed for density
measurement.
Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit.
Different communications tools may use different labels.
Options for Density Measurement UnitTable 4-10:
Unit description
Label
Display ProLink II ProLink III Field Communica-
tor
Specific gravity unit (not tem-
perature-corrected)
SGU SGU SGU SGU
Grams per cubic centimeter
G/CM3 g/cm3 g/cm3 g/Cucm
Grams per liter
G/L g/l g/l g/L
Grams per milliliter
G/mL g/ml g/ml g/mL
Kilograms per liter
KG/L kg/l kg/l kg/L
Kilograms per cubic meter
KG/M3 kg/m3 kg/m3 kg/Cum
Pounds per U.S. gallon
LB/GAL lbs/Usgal lbs/Usgal lb/gal
Pounds per cubic foot
LB/CUF lbs/ft3 lbs/ft3 lb/Cuft
Pounds per cubic inch
LB/CUI lbs/in3 lbs/in3 lb/CuIn
API gravity
D API degAPI degAPI degAPI
Short ton per cubic yard
ST/CUY sT/yd3 sT/yd3 STon/Cuyd
Configure process measurement
46 Micro Motion
®
Model 2700 Transmitters with Analog Outputs